Closed Sampler Aspiration Needle - Abbott CELL-DYN 3000 Operator's Manual

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Maintenance

Closed Sampler Aspiration Needle

9-16
Search
Book TOC
The aspiration needle on the closed sampler version of the CELL-DYN
3000 should be cleaned regularly to remove protein buildup or debris and
reduce the possibility of a blockage.
Materials Required
1. CELL-DYN Enzymatic Cleaner (brought to room temperature)
2. Diluent
3. Three empty Vacutainer® tubes
4. Gloves, lab coat, and suitable eye protection
Procedure
1. Aliquot approximately 2 mL of enzymatic cleaner (undiluted) into
one of the Vacutainer tubes.
2. Aliquot approximately 2 mL of diluent into each of the other two
Vacutainer tubes.
3. If necessary, from the
SAMPLER] to select the closed mode.
4. Press [SPECIMEN TYPE] followed by [BACKGROUND].
5. Aspirate the enzymatic cleaner.
6. Aspirate the Diluent.
7. Check to see that the
5 background counts on the last diluent sample. If the counts are
unacceptable, refer to
8. Discard all tubes used.
Go Back
Data Station RUN
Screen, press [CLOSED
background counts
are acceptable by running
Chapter 10,
Troubleshooting.
CELL-DYN® 3000 System Operator's Manual
Chapter 9
9140240E — May 1995