Chapter 6 Technical Specifications; General Specifications; Performance Specifications; Table 6-6. General Specifications - Abbott PLEX-ID SP User Manual

Table of Contents

Advertisement

specifications
Performance
specifications
Abbott Laboratories
Chapter 6
Technical Specifications
General
The general specifications are presented in Table 6–6.
Table 6–6. General specifications

General specifications

Overall dimensions
– instrument
– transport package
Weight
– instrument
– incl. transport package
Operating conditions
(indoor use)
Transportation conditions
Storage conditions
Mains power supply
Power consumption
Heat dissipation
Computer interface
Internal software
Normal use
The performance specifications are presented in Table 6–7.
Table 6–7. Performance specifications

Performance specifications

Processing volume
Capacity
Collection efficiency of
Continued
ca. 680 mm (W) x 600 mm (D) x 380 mm (H)
[26.8" (W) x 23.6" (D) x 15" (H)]
800 mm (W) x 1200 mm (D) x 710 mm (H) 31.5" (W) x 47.2" (D) x 28" (H)
ca. 28 kg [62 lbs.]
ca. 44 kg [97 lbs.]
+5°C to +40°C; maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 50% relative humidity at 40°C
Indoor use only
-40°C to +60°C, packed in transport packaging
-25°C to +50°C, packed in transport packaging
100 – 240 Vac, 50/60 Hz, nominal
Automatic voltage detection
150 VA max.; 55 VA standby
341 BTU max.
USB or serial RS-232C port. Baud rate 9600.
Character format 1 start bit, 8 data bits, 1 stop bit, and no parity.
Flow control XON/XOFF.
Space for max. 512 internal protocols
10 runs/day, 250 days/year, with 40 min protocols having a medium speed
setting, RT
24: 200 µl – 5 mL
96: 20 – 1000 µl
96 samples/run
> 95%, 96 Well plate, neutral wash buffer containing detergent,