Chapter 5
LAB ID
SETUP
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Search
Book TOC
7. Press [RETURN] to return to the
NOTE: The X-B status is shown on the
DISPLAY SPECIMEN Screen
Box (for example, X-B: 10/IN or X-B: 13/OUT2). The number
of specimens in the current X-B batch is noted (1-19). The status
of the previous batch is displayed following this number. The
status may be OUT, OUT2,
or IN.
• OUT indicates that the mean of the previous batch was outside
the action limits.
• OUT2 indicates that the mean of the last 2 or more completed
batches fell outside the action limits.
• IN indicates that the mean of the previous batch was within the
action limits.
The feature enabling the transfer of QC limits and data into a QC file
from a floppy disk is not currently available. Therefore the following
keys mentioned in this manual should not be used at this time: [LAB ID
SETUP], [LOAD FROM DISK], and [WRITE QC TO DISK].
Operating Instructions
Go Back
QC SETUP MENU
below the Data Station Status
Screen.
RUN Screen
or
5-15