Troubleshooting
CLEAR
FAULTS
RAW DATA
SUMMARY
10-10
Search
Book TOC
864.7
756.6
648.5
540.4
432.4
324.3
216.2
108.1
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5
WBC
RBC
PLT
CNT RATE
CNT RATE
CNT RATE
Figure 10.7:
When the [CLEAR FAULTS] key is pressed, the Analyzer returns to the
READY state if the corrective action taken resolved the problem. If the
corrective action did not correct the problem, the fault status does not
change.
NOTE: Only operator-correctable faults can be cleared with the
[CLEAR FAULTS] key.
When the [RAW DATA SUMMARY] key is pressed, detailed
information pertaining to the last cycle run is displayed. An example of
the
RAW DATA SUMMARY Screen
useful information for the operator, the
reference and sample
The information on metering times may be used to assist in
troubleshooting chronic clog or flow error messages. The HGB reference
and sample readings may be used to assist in troubleshooting erratic or
imprecise HGB results.
Go Back
DIAGNOSTICS MENU
Ready
WBC RATE GRAPH
WBC Count Rate Graph
is shown in
readings, is highlighted in
Chapter 10
Jan 20 1993
12:29
Operator ID
734
Sequence #
2716
PRINT
Figure
10.8. The most
metering times and the HGB
Figure
10.8.
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995