Abbott CELL-DYN 3000 Operator's Manual page 305

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Calibration
Instrument:______________________________________
Date:________________________________________
Operator:________________________________________
________________________________________________________________________________________
_______
Factor
%Diff*
WBC
RBC
HGB
MCV
PLT
* Delete the sign of the Factor % Difference before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
6-50
Search
CELL-DYN 3000
Auto-Cal Calibration Worksheet
Auto-Cal Calibration Criteria
Validation Range
Cal Not Required
<
1.5%
<
1.0%
<
1.0%
<
1.0%
<
3.0%
Book TOC
Go Back
Cal Range
Cal Required
>1.5% but
<
10%
<
>1.0% but
10%
>1.0% but
<
10%
>1.0% but
<
10%
>3.0% but
<
15%
Cal Limit
Do Not Cal**
>10%
>10%
>10%
>10%
>15%
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Chapter 6
Cal?
Y/N

Advertisement

Table of Contents