The Spec ID and Sequence Number for the specimen displayed appears
in the upper left hand corner of the display and printout. The information
in the upper right hand corner displays the current operational status of
the Analyzer. This includes the current date, time, operator ID, current
specimen sequence number and sampler status (Open or Closed).
The data displayed on the left side of the
Screen
and printout provide supplemental information regarding the
suspect population flags for BAND, IG, VAR LYM, VLYM/BLAST and
NRBC. The word SUSPECT will also appear above the suspect
population flags as it does on the
Screens. The number of cells found in each region associated with the
suspect population flag is displayed as a percentage of the total WBC
value. The region percentages displayed on the
DIAGNOSTICS Screen
flags displayed on the
Screen, thus providing additional information as to potential severity of the
flag.
The graphs located on the right side of the
DIAGNOSTICS Screen
been selected for the specimen displayed. The parameter set is indicated
on the
RUN Screen
or
current 2000 specimens in the Data Log, the parameter set can be edited
after the sample has been processed using the [EDIT SPECIMEN] soft
key on the
DATA LOG
NOTE: These data are provided for laboratory use only. The
information may be incorporated into your laboratory's review
criteria and guidelines. Whenever a morphological flag is
displayed, a stained smear should be reviewed.
Go Back
FLAGGING DIAGNOSTICS
RUN
and
DISPLAY SPECIMEN
are used as criteria for the suspect population
RUN Screen
and the
DISPLAY SPECIMEN
FLAGGING
are determined by the parameter set which had
DISPLAY SPECIMEN
Screen.
Operating Instructions
FLAGGING
Screen. On the most
5-57