Table 10.4: Non-Functional Fault Conditions - Abbott CELL-DYN 3000 Operator's Manual

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Troubleshooting
10-74
Search
Book TOC
Table 10.4:
Non-Functional Fault Conditions
Analyzer or Data Station will not power ON
– – – – – Probable Cause – – – –
1.
Power source is defective.
2.
Power cord is not securely
connected to the Analyzer
or is not connected to the
power outlet.
3.
Analyzer fuse is blown
or incorrect.
CAUTION: Always turn the Analyzer power switch OFF and
disconnect the power cord from the receptacle before checking
or replacing any fuse.
4.
Defective power switch or
other system malfunction.
__________________________________________________________
– – – – – Probable Cause – – – –
1.
Data Station power
switch is turned OFF.
2.
Data Station brightness
control is turned down.
Station until the image is
3.
Defective Data Station
or other component.
Go Back
– – – – Corrective Action – – – –
1. Verify that the power
switch is OFF and
connect the system to a
known-good power source.
2. Ensure that the power cord
is securely connected to
the Analyzer or Data
Station and verify that it is
connected to the power
outlet.
3. The Analyzer fuse is located
above the power cord con-
nector on the rear panel.
Check the
in the
Troubleshooting
Procedures
chapter.
4. Call for technical assistance.
No screen display
– – – – Corrective Action – – – –
1. Turn the power switch ON.
2. Adjust the brightness control
on the right side of the Data
visible.
3. Call for technical assistance.
CELL-DYN® 3000 System Operator's Manual
Chapter 10
fuse
as directed
section of this
9140240E — May 1995