Chapter 10
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Search
Book TOC
Table 10.1
Instrument Status Conditions (Continued)
– – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The Analyzer hardware initialization is complete.
Press [PRIME] or [RUN] to initiate the priming cycle and return the
instrument to the READY state. Resume operation when the cycle is
complete.
__________________________________________________________
– – – – – – –– – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The Analyzer was idle for a predetermined time period and therefore the
valves are being exercised to be sure that tubing is not pinched shut.
Resume processing when the READY message is displayed in the status
box.
__________________________________________________________
– – – – – – –– – – – – – – Explanation/Action– – – – – – – – – – – – – – –
A low value has been detected for the WBC and/or PLT count. The
Analyzer is automatically extending the cycle to count more cells.
Press [PRIME] to resume processing.
__________________________________________________________
– – – – – – –– – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The [CLEAR ORIFICE] key was pressed or the instrument
automatically performed the Clear Orifice function in response to a clog
or flow problem.
Wait until the READY message is displayed in the status box and repeat
the sample.
Go Back
Initialized/SB
Unpinching valves/SB
Extended Count/SB
Clearing RBC Orifice/SB
Troubleshooting
10-47