8. When the plate is clean, remove it from the cleaning solution and
thoroughly rinse it with a stream of deionized water. Do not dry the
aperture plate.
9. Insert the plate between the transducer chambers with the
orientation notch on the bottom leading edge. Carefully push the
plate into place. Be sure that it is completely seated between the
chambers.
10. Move the release lever back to the left to securely hold the plate in
place. (See
Figure
11. Press [FILL RBC] to refill the transducer assembly and the
associated tubing.
12. Replace the upper and lower front covers.
13. Press [MAIN] followed by [RUN] to return to the
14. Verify that the background counts are acceptable by running 5
background counts.
15. When an acceptable background count has been obtained, record
the count time in the instrument logbook. This is the baseline count
time for the instrument.
NOTE: The baseline value may only be obtained immediately
after the RBC/PLT aperture has been cleaned.
16. Verify that controls are acceptable. If unacceptable, refer to
Chapter 10,
Troubleshooting.