Abbott CELL-DYN 3000 Operator's Manual page 285

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Calibration
9.______
Verify the Open mode precision by analyzing a fresh, normal whole blood sample 10 times in
succession. Run the sample in an empty control file and record the CVs below or attach a file
printout to this document.
PARAMETER
WBC
RBC
HGB
MCV
PLT
10.______ Verify the Closed mode precision of the Sample Loader by obtaining 15 mL of blood from the
same donor. Aliquot the blood into five 5 mL tubes that contain no anticoagulant. Run each tube
two times in succession. If you have a Closed Sampler, verify its precision by analyzing a fresh,
normal whole blood sample 10 times in succession. Run the samples in an empty control file and
record the CVs below or attach a file printout to this document:
PARAMETER
WBC
RBC
HGB
MCV
PLT
11._____
If any problems are detected during the procedures outlined above, document the problem and the
corrective action on the following page.
6-30
Search
Book TOC
CV% LIMIT
CV
<2.5%
________
<1.7%
________
<1.2%
________
<1.5%
________
<6.0%
________
CV% LIMIT
CV
<2.5%
________
<1.7%
________
<1.2%
________
<1.5%
________
<6.0%
________
Go Back
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Chapter 6