Setting The Address/Addressable Mode - YOKOGAWA WT110 User Manual

Digital power meter
Table of Contents

Advertisement

11.5 Setting the Address/Addressable Mode

Relevant Keys
Operating Procedure
• Perform operations following the thick line in the below menu.
• Press the ENTER key to confirm the selection or setting.
• When you want to leave the current menu during operation, press the key described
1.
SHIFT
Explanation
IM 253401-01E
A
hour
min
SAMPLE
V OVER
A OVER
MODE
B
RMS
V MEAN
DC
C
hour
min
sec
SCALING
AVG
FILTER
STORE
* Shows the operation panel of the WT130. For the differences
between WT110 and WT130, refer to section 2.2, page 2-2, 2-3
under step 1. The confirmed settings made until that point will be kept.
Setting the mode
( Display C )
LOCAL
2.
INTERFACE
Mode Setting
Refer to page 11-1 for details.
Address Setting
A particular address is assigned to each device connected to the GP-IB interface so that each
device can be recognized by every device. Therefore, an address must be assigned to this
instrument when it is connected to a personal computer.
Address setting range: 0 to 30
The initial value is "1". Initializating the instrument will not result in changing the address
setting.
Talk-only Function
This function only allows the instrument to send data to other devices. If talk-only is off, the
instrument can both send and receive data. In talk-only mode, the instrument cannot be
controlled by the controller.
Terminator
When this instrument is used as a listener
Use "CR+LF", "LF" or "EOI" as the receiving terminator.
When this instrument is used as a talker
The sending terminator is set using the DL command. The initial setting is "CR+LF+EOI".
Note
• It is not possible for this instrument to receive data if the "CR" terminator is sent from the controller. It is
also not possible to set "CR" as the terminator which is to be sent from this instrument.
m
V
VA
1
2
3
AUTO
AUTO
var
V RANGE A RANGE
k
A
FUNCTION
ELEMENT
MODE
M W
TIME
m
V
PF
1
2
3
k
A
deg
FUNCTION
ELEMENT
M W
%
START
STOP
HARMONICS MEMORY
REMOTE
m
V
Hz
1
2
3
LOCAL
SETUP
k
A
h
FUNCTION
ELEMENT
INTERFACE
OUTPUT
M W
h
1Φ3W
3Φ3W
RECALL
HARMONICS
3Φ4W
3V3A
Setting the address
3.
( Display C )
ENTER
4.
up down
5.
SHIFT
Setting the interval
( Display C )
3.
hour min sec
ENTER
4.
Same as step 4 to 5
5.
mentioned above.
(Select this when setting commands according to IEEE 488.2-1987.
Note that this menu only appears in case of version 1.11 and later.
Refer to page 3-14 to confirm your version.)
Displays
HOLD
relevant
TRIG
keys and
ENTER
indicators
INTEGRATOR
RESET
INTEG SET
SHIFT
WIRING
6.
ENTER
End
cursor shift
6.
ENTER
End
11-9
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt130

Table of Contents