14.4 Error Codes and Corrective Actions

Error Codes for Operation and Measurement
Error Code
11
12
13
14
15
16
17
18
19
30
31
32
33
41
42
43
44
45
46
47
50
51
52
53
IM 253401-01E
Description
Received a command not used by this
instrument.
Parameter value specified is outside the
allowed range.
Attempted to execute a key operation or
received a communications command, while
integration was running or was interrupted,
that cannot be executed or received in such
a state.
Attempted to set auto range while external
sensor range is selected.
Attempted to execute a command or key
operation that was protected.
Attempted to execute a key operation or
received a communications command, while
harmonic analysis was being performed or
was interrupted, that cannot be executed or
received in such a state.
Print output time-out.
No data to be printed or not in the printing
mode.
Attempted to execute a key operation or
received a communications command, while
storing/recalling of data being performed.
Invalid file data.
File is damaged.
No data stored in the internal memory.
No space to store data in the internal
memory.
⋅ Attempted to start integration while there is
an overflow condition.
⋅ Attempted to start integration after
integration time has reached its preset
value.
Attempted to start integration while
integration is in progress.
Measurement stopped due to overflow
during integration or due to a power failure.
Attempted to stop integration even though
integration was not in progress.
Attempted to reset integration even though
integration was not in progress or integration
mode was not selected.
Attempted to start integration while
measurement of peak overflow was in
progress or during an overrange condition.
Attempted to start integration in continuous
integration mode when integration preset
time was set to "0".
A/D conversion time-out.
Measurement data overflow occurred.
"oL" is displayed.
Voltage peak overflow occurred.
V OVER indicator lights up.
Current peak overflow occurred.
A OVER indicator lights up.
Corrective Action
Check for error in the command
sent.
Correct the value.
Check whether integration is in
progress or is interrupted.
It is not possible to set auto
range while external sensor
range is selected.
Check whether the command or
key operation is correct.
Check whether harmonic
analysis is in progress or is
interrupted.
Store data in the internal
memory or select the proper file
to be stored.
Reset integration.
Set a correct preset time.
Reference page
11-12
App. 2
7-8, 11-12
4-5
-
8-5
10-18
10-17
-
-
-
Ch. 9
9-2
7-5
7-8
7-8
7-8
7-6
2-4
14
7-4
-
2-4
2-4
2-4
14-11