Appendix 2.3 Commands
Output Format/Data Format of Normal Measured/Computed Data and Harmonic Analysis Data
Data Format of Normal Measured/Computed Data
Output Format of Normal Measured/Computed Data
Output Example of Normal Measured/Computed Data
App2-24
The output format/data format of normal measured/computed data and harmonic analysis data
which is requested by MEASure[:NORMal]:VALue? or MEASure:HARMonics:VALue?, is as
follows.
• All data of the <harmonic analysis function> are output in the <NR3> format.
(Example) 99.99E+00
V,A,W,VA,VAR,PF,DEGR,VHZ,AHZ,VPK,APK,MATH →mantissa: max. 4 digits + exponent: 2 digits
WH,WHP,WHM,AH,AHP,AHM
• The sign of the mantissa will only be applied in case of negative values. However, phase lead
and lag (in case of phase angle (DEG)) will be shown as follows.
→ +180.0E+00
LEAD
→ -180.0E+00
LAG
→ 0.0E+00 (The mantissa will be proceeded by a space)
in phase
• In case of overrange or computation over,"9.9E+37"(+∞) will be output.
(i.e. in case the display shows –oL–, –oF–, PFErr, dEGEr, ErrLo, or ErrHi)
• In case no data is present (i.e. the display shows – – – – –), "9.91E+37" (NAN) will be output.
• The integration elasped time is output as hours, minutes, seconds in the <NR1> format.
(Example) 999,59,59
The communication output is set ON by any of the commands starting with
"MEASure[:NORMal]:ITEM" and the normal measured/computed data or integration elasped time
are output according to the following order of priority. Besides, in case of recalling normal
measurement or integration data, the data number will be output in <NR1> format as well. Data
will be output in the following order corresponding to each element. However, note that for model
253401 only element 1 is valid, and for model 253502 only element 1, 3 and Σ are valid.
(0. Data number in case of recalling)
→V2
1. V1
→A2
2. A1
→W2
3. W1
→VA2
4. VA1
→VAR2
5. VAR1
→PF2
6. PF1
→DEGR2
7. DEGR1
→VHZ2
8. VHZ1
→AHZ2
9. AHZ1
→WH2
10. WH1
→WHP2
11. WHP1
→WHM2
12. WHM1
→AH2
13. AH1
→AHP2
14. AHP1
→AHM2
15. AHM1
16. TIME (integration elasped time)
Each data is divided by a comma"," and is ended by the terminator <RMT>.
• Output example for model 253502 after having sent the following commands.
(Sent)
MEASURE:NORMAL:ITEM:PRESET NORMAL
MEASURE:NORMAL:VALUE?
(Received data) 10.04E+00,10.02E+00,10.03E+00,49.41E+00,49.52E+00,
49.47E+00,429.0E+00,429.2E+00,0.858E+03
(Data contents)
V1:10.04E+00
A1:49.41E+00
W1:429.0E+00
→mantissa: max. 6 digits + exponent: 2 digits
(max. 5 digits in case of negative value)
→V3
→V∑
'
→A3
→A∑
→W3
→W∑
→VA3
→VA∑
→VAR3
→VAR∑
→PF3
→PF∑
→DEGR3
→DEGR∑
→VHZ3
→VHZ∑
→AHZ3
→AHZ∑
→WH3
→WH∑
→WHP3
→WHP∑
→WHM3
→WHM∑
→AH3
→AH∑
→AHP3
→AHP∑
→AHM3
→AHM∑
V3:10.02E+00
A3:49.52E+00
W3:429.2E+00
V∑:10.03E+00
A∑:49.47E+00
W∑:0.858E+03
IM 253401-01E