Harmonics Group - YOKOGAWA WT110 User Manual

Digital power meter
Table of Contents

Advertisement

Appendix 2.3 Commands
2.3.6

HARMonics Group

The commands in the HARMonics group relate to the harmonic analysis function. This allow you to make the same settings and
inquiries as when using the HARMONICS key on the front panel and the corresponding menus. This group is only useful in case
your instrument is equipped with the /HRM option.
:HARMonics
:
HARMonics?
Function
Queries all settings relating to harmonic analysis.
Syntax
HARMonics?
Example
HARMONICS?→:HARMONICS:STATE 0;ELEMENT 1;
SYNCHRONIZE V,1;THD IEC;DISPLAY:ORDER 1
HARMonics:DISPlay?
Function
Queries all settings concerning the display in case of
harmonic analysis.
Syntax
HARMonics:DISPlay?
Example
HARMONICS:DISPLAY?
HARMONICS:DISPLAY?→:HARMONICS:
DISPLAY:ORDER 1
HARMonics:DISPlay:ORDer
Function
Sets the order of the harmonic component to be shown
on display B, queries the current setting.
Syntax
HARMonics:DISPlay:ORDer {<NRf>}
HARMonics:DISPlay:ORDer?
{<NRf>}=1 to 50
Example
HARMONICS:DISPLAY:ORDER 1
HARMONICS:DISPLAY:ORDER?→:HARMONICS:
DISPLAY:ORDER 1
HARMonics:ELEMent
Function
Sets the element for harmonic analysis/queries the
current setting.
Syntax
HARMonics:ELEMent {<NRf>}
HARMonics:ELEMent?
{<NRf>}= 1
1, 3 (WT130 three-phase, three-wire
1 to 3(WT130 three-phase, four-wire
Example
HARMONICS:ELEMENT 1
HARMONICS:ELEMENT?→:HARMONICS:ELEMENT 1
App2-18
;
STAT e
<Space>
?
ELEMent
<Space>
?
SYNChroniz e
<Space>
THD
<Space>
?
DISPlay
:
ORDer
(WT110 single-phase model)
model)
model)
OFF
ON
<NRf>
<NRf>
V
,
A
?
IEC
CSA
;
<Space>
?
?
?
HARMonics[:STATe]
Function
Sets the harmonic analysis mode ON/OFF, queries the
current setting.
Syntax
HARMonics[:STATe] {<Boolean>}
HARMonics[:STATe]?
Example
HARMONICS:STATE ON
HARMONICS:STATE?→:HARMONICS:STATE 1
HARMonics:SYNChronize
Function
Sets the fundamental frequency for PLL
synchronization (PLL source)/queries the current
setting.
Syntax
HARMonics:SYNChronize
{(V|A),(<NRf>|ELEMent<1-3>)}
HARMonics:SYNChronize?
Example
HARMONICS:SYNCHRONIZE V,1
HARMONICS:SYNCHRONIZE?→:HARMONICS:
SYNCHRONIZE V,1
HARMonics:THD
Function
Sets the computation method for harmonic distortion
(THD) for harmonic analysis/queries the current
setting.
Syntax
HARMonics:THD {IEC|CSA}
HARMonics:THD?
Example
HARMONICS:THD IEC
HARMONICS:THD?→:HARMONICS:THD IEC
<NRf>
ELEMent
<x>
<NRf>
IM 253401-01E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt130