11.4 Output Format for Normal Measured/Computed
Data, Harmonic Analysis Data, Set-up

Parameters and Error Codes

Output Format of Normal Measured/Computed Data
11-4
Data Format
Measured data normally consists of a 6-byte header and 11 bytes of data
Header
Data
Header Section
The header section consists of 6 bytes (h1 to h6).
h1
h2
h3
h4
h1 to h3: data type
V_ _
: voltage
VA_
: Apparent power
HzV
: Voltage frequency
Ah_
: Ampere hour
Apk
: Peak current value
CV2
: V2 crest factor
CA2
: A2 crest factor
A+B
: (display A)+(display B)
A * B
: (display A) * (display B)
Wh+
: Positive watt hour
Ah+
: Positive ampere hour
HMS : Elapsed time of integration
(Efficiency, crest factor, and the results of four arithmetic operations can be output on
WT110/WT130 with ROM version 2.01 and later. However, different instruments have
different restrictions on the output of these parameters.)
h4: Element
1: Element 1
2: Element 2
h1-h4: data type (only when the computing equation is A/B
ROM version 2.01 or later)
2
A/B
: (display A)/(display B)
2
2
A
/B: (display A)
/(display B)
h5: Data state
N: normal
I: Overrange
E: No data
h6: Indicates data lag/lead in case of DEG data type. In case of other data types, _ (space) will
occur.
G: Lag
D: Lead
Data Section
The data section consists of 11 bytes.
d1
d2
d3
d4
d1
: polarity; _ (space) or – (minus)
d2 to d8
: mantissa, floating-point number of the maximum six digits
d9 to d11 : exponent; E-3→m, E+0, E+3 → k, E+6 → M
Data state in case of an overrange ("oL" is being displayed)
h1
h2
h3
h4
I
Data state in case of a computation overflow
("oF", "PFErr", "dEGEr", "ErrLo", "ErrHi" is being displayed)
h1
h2
h3
h4
O
Data state in case of no data (when the display is - - - - -)
"I" becomes "E" for data during overrange.
Elapsed time of integration
H
M
S
_
d1 to d3 : Hour
d4 : ":"
d5 to d6 : Minute
d7 : ":"
d8 to d9 : Second
h5
h6
A_ _
: Current
Var
: Reactive power
HzA
: Current frequency
DEG : Phase angle
EFF
: Efficiency
CV3
: V3 crest factor
CA3
: A3 crest factor
A–B
: (display A)–(display B)
A/B
: (display A)/(display B)
Wh–
: Negative watt-hour
Ah–
: Negative ampere hour
MEM : Data number in case of recalling
3: Element 3
2
O: Computation overflow
_: Not detectable
d5
d6
d7
d8
d9
_
_
9
9
9
9
_
_
8
8
8
8
_
_
d1
d2
d3
d4
W_ _ : Active power
PF_
: Power factor
Wh_
: Watt hour
Vpk
: Peak voltage value
CV1
: V1 crest factor
CA1
: A1 crest factor
4: ∑
2
2
or A
/B on WT110/WT130 with
P: Peak overflow
d10
d11
9
9
.
E
+
3
8
8
.
E
+
0
d5
d6
d7
d8
d9
IM 253401-01E