12.3 Setting the Mode, Handshaking Method, Data Format and Baud Rate
Circuit idle
state
12-6
Data Format
The RS-232-C interface of this instrument performs communications using start-stop
synchronization. In start-stop synchronization, one character is transmitted at a time. Each
character consists of a start bit, data bits, a parity bit, and a stop bit. Refer to the figure below.
Start bit
The table below shows the data format combinations supported.
Preset value
Start bit
0
1
1
1
2
1
3
1
Baud Rate
The baud rate can be selected from 75, 150, 300, 600, 1200, 2400, 4800 or 9600.
About the Terminator
Data can be received with either "CR+LF" or "LF" terminator. For transmission terminator,
you can select from "CR+LF," "LF," and "CR."
Interval
In case of the talk-only mode, this setting specifies the interval to send data.
Setting range : 00.00.00 (0hr 0min 0sec) to 99.59.59 (99 hrs 59min 59sec)
Initial value
: 00.00.00
Note
The error code 390 may appear depending on the status of this instrument. In such a case, lower the baud
rate.
1 character
Data bit
(7 or 8 bits)
Parity bit
Even, odd or none
Data length
8
7
7
7
Level returns to idle
state (dotted line) or
the start bit the next
data (solid line)
Stop bit
1
1 or 2 bits
2
Parity
Stop bit
No
1
Odd
1
Even
1
No
2
IM 253401-01E