YOKOGAWA WT110 User Manual page 150

Digital power meter
Table of Contents

Advertisement

• While recalling or storing is in progress,
execution error 19 will occur.
SC/SC?
Determines whether or not to use the
scaling function/inquires about the
current setting.
Syntax
SC m <terminator>
"m" indicates whether scaling is ON or OFF.
m=0 : OFF
1 : ON
Query
SC? <terminator>
Example
SC1
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SL
Recalls set-up parameters from a
selected file.
Syntax
SL m <terminator>
"m" indicates file no., and must be set within the
following range.
1 ≤ m ≤ 4
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
• It is not possible to recall communications-
related information (communication mode,
address etc.) using this command.
• While recalling or storing is in progress,
execution error 19 will occur.
SO/SO?
Sets the store function ON/OFF or
inquires about the current setting.
Syntax
SO m <terminator>
"m" indicates whether storage is ON or OFF.
m=0 : OFF
1 : ON
Query
SO? <terminator>
Example
SO1
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SR/SR?
Sets the storage interval/inquires about
the current setting.
Syntax
SR m1,m2,m3 <terminator>
"m1" indicates the hour
0 ≤ m1 ≤ 99
"m2" indicates the minutes
0 ≤ m2 ≤ 59
"m3" indicates the seconds
0 ≤ m3 ≤ 59
Query
SR? <terminator>
Example
SR0,0,0
Description
• Parameter error 12 will occur if an illegal value
is set.
• When the storage interval is set to 0hrs, 0min,
0sec, refer to page 9-2.
• While recalling or storing is in progress,
execution error 19 will occur.
SS
Stores set-up parameters into a selected
file.
Syntax
SS m <terminator>
"m" indicates file no., and must be set within the
following range.
1 ≤ m ≤ 4
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
• The following set-up parameters can be stored:
All set-up parameters which can be output by
the OS command
Information related to communications (GP-
IB, RS-232-C etc.)
IM 253401-01E
• While recalling or storing is in progress,
execution error 19 will occur.
TM/TM?
Sets integration preset time/inquires
about the current setting.
Syntax
TM m1,m2 <terminator>
"m1" indicates hour, and must be set within the
following range.
0 ≤ m1 ≤ 999
"m2" indicates minute, and must be set within the
following range.
0 ≤ m2 ≤ 59
Query
TM? <terminator>
Example
TM0,0
Description
• Parameter error 12 will occur if an illegal value
is set.
• While recalling or storing is in progress,
execution error 19 will occur.
WR/WR?
Sets the wiring system/inquires about the
current setting.
Syntax
WR m <terminator>
m=2 :1Φ3W
3 :3Φ3W
4 :3Φ4W (available only for the 253503)
5 :3V3A (available only for the 253503)
Query
WR? <terminator>
Example
WR2
Description
• Parameter error 12 will occur if an illegal value
is set.
YC/YC?
Sets the display channel while the
comparator function is ON/inquires about
the current setting.
Syntax
YC m <terminator>
"m" indicates the channel number for display
• in case of single mode
m=1 :Displays limit and measurement value
on display 1
2 :Displays limit and measurement value
on display 2
3 :Displays limit and measurement value
on display 3
4 :Displays limit and measurement value
on display 4
• in case of dual mode
m=1,2 :Displays limit and measurement value
on display 1 and 2 respectively
m=3,4 :Displays limit and measurement value
on display 3 and 4 respectively
Query
YC? <terminator>
Example
YC1
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
YM/YM?
Sets the mode of the comparator function/
inquires about the current setting.
Syntax
YM m <terminator>
"m" indicates the display mode
m=0 :Single mode
1 :Dual mode
Query
YM? <terminator>
Example
YM1
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
YO/YO?
Sets the comparator function ON/OFF or
inquires about the current setting.
Syntax
YO m <terminator>
"m" indicates whether the comparator function is
ON/OFF
m=0 :OFF
1 :ON
Query
YO? <terminator>
Example
YO1
Description
• Parameter error 12 will occur if "m" is set to an
illegal value.
Appendix 1.1 Commands
App1-9
App

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt130