YOKOGAWA WT110 User Manual page 172

Digital power meter
Table of Contents

Advertisement

CONFigure?
Function
Queries all the settings related to the measurement
conditions.
Syntax
CONFigure?
Example
CONFIGURE?→:CONFIGURE:WIRING P1W3;MODE RMS;
VOLTAGE:RANGE 600.0E+00;AUTO 1;:CONFIGURE:
CURRENT:RANGE 20.0E+00;AUTO 1;ESCALING:
ELEMENT1 50.00E+00;ELEMENT2 50.00E+00;
ELEMENT3 50.00E+00;:CONFIGURE:FILTER 0;
SCALING:STATE 0;PT:ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;ELEMENT3 1.000E+00;
:CONFIGURE:SCALING:CT:ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;ELEMENT3 1.000E+00;
:CONFIGURE:SCALING:SFACTOR:ELEMENT1 1.000E+00
;ELEMENT2 1.000E+00;ELEMENT3 1.000E+00;
:CONFIGURE:AVERAGING:STATE 0;TYPE LINEAR,8
[CONFigure]:AVERaging?
Function
Queries all the setting values related to the averaging
function.
Syntax
[CONFigure]:AVERaging?
Example
[CONFIGURE]:AVERAGING?→:CONFIGURE:AVERAGING:
STATE 0;TYPE LINEAR,8
[CONFigure]:AVERaging[:STATe]
Function
Sets averaging ON/OFF, or queries the current status.
Syntax
[CONFigure]:AVERaging[:STATe] {<Boolean>}
[CONFigure]:AVERaging:STATe?
Example
[CONFIGURE]:AVERAGING:STATE OFF
[CONFIGURE]:AVERAGING:STATE?→:CONFIGURE:
AVERAGING:STATE 0
[CONFigure]:AVERaging:TYPE
Function
Sets the averaging type and constant, queries the
current setting.
Syntax
[CONFigure]:AVERaging:TYPE {LINear|
EXPonent},{<NRf>}
[CONFigure]:AVERaging:TYPE?
{<NRf>}=8, 16, 32, 64 (averaging constant)
Example
[CONFIGURE]:AVERAGING:TYPE LINEAR,8
[CONFIGURE]:AVERAGING:TYPE?→:CONFIGURE:
AVERAGING:TYPE LINEAR,8
[CONFigure]:CURRent?
Function
Queries all setting values relating to the current range
(external sensor range)
Syntax
[CONFigure]:CURRent?
Example
[CONFIGURE]:CURRENT?→:CONFIGURE:CURRENT:
RANGE 20.0E+00;AUTO 1;ESCALING:
ELEMENT1 50.00E+00;ELEMENT2 50.00E+00;
ELEMENT3 50.00E+00
[CONFigure]:CURRent:AUTO
Function
Sets the current auto range ON/OFF, or queries the
current setting.
Syntax
[CONFigure]:CURRent:AUTO {<Boolean>}
[CONFigure]:CURRent:AUTO?
Example
[CONFIGURE]:CURRENT:AUTO ON
[CONFIGURE]:CURRENT:AUTO?→:CONFIGURE:
CURRENT:AUTO 1
[CONFigure]:CURRent:ESCaling?
Function
Queries all scaling values for the external sensor.
Syntax
[CONFigure]:CURRent:ESCaling?
Example
[CONFIGURE]:CURRENT:ESCALING?→:CONFIGURE:
CURRENT:ESCALING:ELEMENT1 50.00E+00;
ELEMENT2 50.00E+00;ELEMENT3 50.00E+00
IM 253401-01E
[CONFigure]:CURRent:ESCaling[:ALL]
Function
Sets the scaling values for the external sensor for all
elements at once.
Syntax
[CONFigure]:CURRent:ESCaling[:ALL] {<NRf>}
{<NRf>}= 0.001 to 1000
Example
[CONFIGURE]:CURRENT:ESCALING:ALL 50.00
Description
The setting values differ as follows.
Less than 1.000 : Three digits after the floating-point
are valid.
1.000 to 1000 : The first five digits are valid.
[CONFigure]:CURRent:ESCaling:ELEMent<x>
Function
Sets the scaling values for the external sensor for each
element separately, queries the current setting.
Syntax
[CONFigure]:CURRent:ESCaling:
ELEMent<x> {<NRf>}
[CONFigure]:CURRent:ESCaling:ELEMent<x>?
<x>=1
(WT110 single model)
1, 3
(WT130 three-phase, three-wire
model)
1 to 3 (WT130 three-phase, four-wire
model)
{<NRf>}=0.001 to 1000
Example
[CONFIGURE]:CURRENT:ESCALING:ELEMENT1 50.00
[CONFIGURE]:CURRENT:ESCALING:ELEMENT1?→
:CONFIGURE:CURRENT:ESCALING:
ELEMENT1 50.00E+00
Description
Setting values differ as described at [CONFigure]:
CURRent:ESCaling[:ALL].
[CONFigure]:CURRent:RANGe
Function
Sets the current range (external sensor input range),
queries the current setting.
Syntax
[CONFigure]:CURRent:RANGe
{<current>|(EXTernal,<voltage>)}
[CONFigure]:CURRent:RANGe?
<current>=500mA to 20A (0.5, 1, 2, 5, 10,
20A)
<voltage>=50mV to 200mV (50, 100, 200mV, for
/EX2 option)
=2.5V to 10V
EX1 option)
Example
Setting of current range/query
[CONFIGURE]:CURRENT:RANGE 20A
[CONFIGURE]:CURRENT:RANGE?→:CONFIGURE:
CURRENT:RANGE 20.0E+00
Setting of external sensor input range/
query (in case of /EX2 option)
[CONFIGURE]:CURRENT:RANGE EXTERNAL,50MV
[CONFIGURE]:CURRENT:RANGE?→:CONFIGURE:
CURRENT:RANGE EXTERNAL,50.0E-03
[CONFigure]:FILTer
Function
Sets the filter ON/OFF, queries the current setting.
Syntax
[CONFigure]:FILTer {<Boolean>}
[CONFigure]:FILTer?
Example
[CONFIGURE]:FILTER OFF
[CONFIGURE]:FILTER?→:CONFIGURE:FILTER 0
[CONFigure]:MODE
Function
Sets the measurement mode of current and voltage,
queries the current setting.
Syntax
[CONFigure]:MODE {RMS|VMEan|DC}
[CONFigure]:MODE?
Example
[CONFIGURE]:MODE RMS
[CONFIGURE]:MODE?→:CONFIGURE:
MODE RMS
Appendix 2.3 Commands
(2.5, 5, 10V, for /
App2-15
App