Turning The Filter On/Off - YOKOGAWA WT110 User Manual

Digital power meter
Table of Contents

Advertisement

4.2

Turning the Filter ON/OFF

Relevant Keys
Operating Procedure
Explanation
IM 253401-01E
A
hour
min
SAMPLE
V OVER
A OVER
MODE
B
RMS
V MEAN
DC
C
hour
min
sec
SCALING
AVG
FILTER
STORE
*Shows the operation panel of the WT130. For the differences
between WT110 and WT130, refer to section 2.2, page 2-2, 2-3
• Perform operations following the thick line in the below menu.
• Press the ENTER key to confirm the selection or setting.
• When you want to leave the current menu during operation, press the key described
under step 1. The confirmed settings made until that point will be kept.
Selecting the filter
1.
(Display C)
SETUP
2.
Filter Function
The instrument will perform measurements after synchronizing to the cycle of the input signal.
Consequently, the frequency of the input signal can be measured properly. The filter, at a cut-
off frequency of 300Hz, will only be applied to the frequency measurement circuit and will
remove noise from distorted and inverted waves, etc.. This allows the frequency to be measured
correctly which improves the accuracy of each measurement value. The filter will not be applied
to the voltage and current circuit. The initial value is OFF.
Note
The filter setting cannot be changed while integration is being carried out.
m
V
VA
1
2
3
AUTO
AUTO
var
V RANGE A RANGE
k
A
FUNCTION
ELEMENT
MODE
M W
TIME
m
V
PF
1
2
3
k
A
deg
FUNCTION
ELEMENT
M W
%
START
STOP
HARMONICS MEMORY
REMOTE
m
V
Hz
1
2
3
LOCAL
SETUP
k
A
h
FUNCTION
ELEMENT
INTERFACE
OUTPUT
M W
h
1Φ3W
3Φ3W
RECALL
HARMONICS
3Φ4W
3V3A
(Display C)
3.
5
ENTER
4.
Displays
HOLD
relevant
TRIG
keys and
ENTER
indicator
INTEGRATOR
RESET
INTEG SET
SHIFT
WIRING
.
ENTER
End
4-3