1.5 Setting Addressable Mode

Before You Begin
Operationg Procedure
IM 253710-11E
When you make settings which can be made using the front panel keys of the instrument
or when you output set-up data or waveform data using the controller, the following
settings must be made.
Setting the address
This function allows you to set the instrument's address for addressable mode within the
range of 0 to 30. Each item of equipment connected via a GP-IB interface has its own
address, by which it can be identified. Care must be taken to ensure that all
interconnected devices are assigned unique addresses.
Note
Do not change the address while the GP-IB interface is being used by the controller.
1.
Press the MISC key.
2.
Press the "GP-IB/RS232" soft key.
3.
Press the "Comm Device" soft key to select "GPIB."
4.
Turn the jog shuttle to set the desired address.
Step 1, 2
Step 4
Step 3