Exchange The Coolant Hose From The Handpiece Or Angle Piece; Exchange The Coolant Hose In The Motor Hose - KaVo INTRAsurg 300 Instructions For Use Manual

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.2 Exchange coolant hose
Coolant hose ④(obtainable approx. 150 mm long under Mat. no. 0.593.0252 and
as a section under Mat. no. 0.065.5188)

7.2.1 Exchange the coolant hose from the handpiece or angle piece.

▶ Pull off the coolant hose ④ Mat. no. 0.593.0252 (150 mm), or
Mat. no. 1.001.3436 (70 mm), or Mat. no. 0.065.5188 (cut section) off the cou‐
pling ③ Mat. no. 1.001.6462 and the handpiece or angle piece and exchange.

7.2.2 Exchange the coolant hose in the motor hose

Note
Run the new hose straight in the direction of tension so that it can be pulled in as
easily as possible. Talcum powder can be used as a lubricant.
▶ Remove the coolant hose ① from the plug-in nipple ② and coupling ③.
58/66