KaVo INTRAsurg 300 Instructions For Use Manual page 19

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
3 Product description | 3.5 Technical data for the 300/300 plus
Input voltage
Input frequency
Power consumption
True power
Speed
Pump delivery rate
delivery pressure
Multifunctional foot control
Only KaVo INTRAsurg 300 Plus:
Multifunctional foot control, identifica‐
tion: blue FBE stirrup, rating plate. Per‐
missible in surgery rooms.
Standard-length hose for the multifunc‐
tional foot control
Special-length hose for the multifunc‐
tional foot control
Only available with the INTRAsurg
300plus.
Standard motor hose length
Mat. no. 1.001.2651
Special length motor hose
Mat. no. 1.004.6825
Mode
continuous operation duty type
The torque precision of the KaVo angle
piece 27:1 (TRA C09 + C3) ranges from
20 to 30 Ncm at 20 to 50 rpm.
Greater deviations are possible with oth‐
er angle pieces.
Transportation and storage conditions
Ambient temperature
Relative humidity
Air pressure
100 V∼/120 V∼/230 V∼
50/60 Hz
Max. 260 VA
Approximately 7 Watt
300 – 40000 rpm
25 – 150 ml/min
1.5 – 2.2 bar
Protection class IPX1
Protection class IPX8
1.90 m
3.5 m
2.00 m
3.0 m
30 sec. operation/ 9 min.sec pause
± 3 Ncm
-20℃ to +70℃
5% to 95%
700 hPa to 1060 hPa
17/66