KaVo INTRAsurg 300 Instructions For Use Manual page 57

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
Malfunction
Display:
>>remark 65
-motor is overloaded
Display:
>>warning 50
-motor is overloaded
Display:
>>remark 66
-motor not plugged
Display:
>>remark 67
-motor not plugged.
Display:
>>warning 36
-missing foot control
Display:
error 3
Display:
error 4
Display:
error 29
Display:
warning 49
Cause
The hand and angle piece is im‐
properly attached.
The lamp is faulty.
The catch is dirty.
No suitable light, hand and angle
piece
Motor overload warning
Motor is overloaded, the maximum
permissible motor output was ex‐
ceeded.
The motor hose is not correctly in‐
serted.
The motor is not correctly connec‐
ted to the motor hose.
When the motor starts, the discon‐
nection of the motor from the hose
is detected.
Defective footswitch or defective
connection to the footswitch, possi‐
bly by unintentional penetration of
moisture.
Internal system error
Internal system error
The motor hose is damaged.
The power consumption of the sur‐
gical motor was too high.
55/66
Remedy
▶ Mount the handpiece and angle
piece until the catch audibly
locks.
▶ Replace the lamp
Mat. no. 1.002.2928.
See also: Instructions for use for the
motor.
▶ Clean the catch (gold contacts).
See also: Instructions for use for the
motor and instrument.
▶ Use a suitable light, hand and
angle piece.
▶ Stop treatment as soon as pos‐
sible and let the motor cool off
for a few minutes.
▶ The motor is blocked for 10 mi‐
nutes; let it cool.
▶ Properly insert the motor hose.
▶ Properly plug in the motor.
▶ Plug the hose onto the motor
and screw it tight. This display
disappears after eight seconds.
The motor can be restarted after
this time.
▶ Dry the footswitch.
Turn the machine off and on
▶ Exchange the pump hose. Con‐
tact service.
▶ Exchange the pump hose. Con‐
tact service.
▶ Check the motor hose and ex‐
change it if necessary
▶ Press the ENTER button and
continue.
Exchange the motor hose.
Exchange surgical motor.