4 First use | 4.5 Connect the surgery motor and coolant reservoir
8
9
4
ESD symbol
CAUTION
Electrostatic charge.
Destruction of electronic components.
▶ Before contacting the hose coupling, touch the bottle holder with your hand to
release static electricity.
▶ Insert the hose coupling ② into the device connection until it audibly locks into
place and the marking ring on the plug is covered.
▶ Shove the coolant hose ④ onto the plug-in nipple ③ by twisting slightly.
▶ Open the pump mechanics by lifting the lock ⑧.
Note
Insert the pump hose into the pump so that it is not clamped or pinched by the lock.
Run all hoses relaxed without tension. Shove the hoses firmly on the plug-in nipple
to ensure a tight connection.
▶ Close the lock ⑧ in the direction of the arrow.
▶ Insert the insertion needle ⑨ into the coolant reservoir, and hang the coolant
reservoir on the bottle holder. Check the seal and seat of the insertion tip. Pre‐
vent fluid from exiting the device.