Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
▶ To leave the display, press the enter button.
All saved torques are deleted.
Note
If you leave the program buy pressing the plus button or the right button on the
multifunctional foot control, all torques are deleted. If you accidently go too far and
skip the torque display, the values can be reactivated and displayed by pressing
the right button on the multifunctional foot control.

5.5 Functions exclusively with the 300 plus program version

Additional functions of the INTRAsurg 300 plus with an INTRA LUX SL550 motor:
▪ Light function.
▪ IPX8 multifunction foot control (blue stirrup, rating plate).
▪ Recognition of the INTRA CL light, hand and angle piece. The transfer ratio and
associated values are shown on the display.
▪ The user can perform additional calibration with the INTRAsurg Calibration
1.002.3569. (27:1 angle piece is required)
▪ Display of the maximum torques. The maximum torque is saved in each step.
The max. torque can be read on the display during surgery or after the program
ends (only with clockwise rotation).
▪ Read-out table for the torque values at the end of applications.

5.5.1 Recognition function

Note
To prevent misinformation, the contact's provided for recognition always need to
be regularly cleaned on the motor and instrument pusher. The contacts always
must be shiny. Wipe off the contacts with a cloth soaked in disinfectant and rub dry.