Drying; Service, Inspection And Testing After Preparation; Sterilisation In A Steriliser In Compliance With Din En 13060 - KaVo INTRAsurg 300 Instructions For Use Manual

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
Permissible disinfectants (the uses correspond to the existing manufacturer's in‐
structions and national guidelines. observe the safety data sheets.) KaVo recom‐
mends the following products based on material compatibility. The microbiological
efficacy must be ensured by the disinfectant manufacturer.
▪ Incidin Liquid (Ecolab)
▪ FD 322 Dürr
▪ Schülke & Mayr Microcide
▪ Isopropanol 70%

6.2.3 Drying

▶ After disinfection and sterilisation: Let the surfaces and plug-in parts for the ho‐
ses and cables (contacts) completely dry in the room before they are used again.

6.2.4 Service, inspection and testing after preparation

▶ Test of cleanliness and intactness, care, repair: check the hose connections for
ease of plugging and secure attachment. Check the function of device settings
and motor operation. Check hose pump for sufficient coolant. Check secure at‐
tachment of hose connections. Check control commands on foot control. Check
recognition function of INTRAsurg 300 plus. Comply with hygiene requirements
(sterility) during the check. If there are any sites of breakage or clearly recog‐
nised changes on the surface, the parts must be checked by the Service.

6.3 Sterilisation in a steriliser in compliance with DIN EN 13060

CAUTION
Product damage due to improper sterilisation.
Damage to the sterile product.
▶ No hot air sterilisation, no chemical cold sterilisation, do not sterilise with eth‐
ylene oxide!
CAUTION
Product damage due to improper sterilisation.
Damage to the sterile product.
▶ An initial vacuum can shorten the product life.
CAUTION
Moisture
Non-sterility
▶ Ensure dryness. Autoclaves with a subsequent vacuum ensure dryness. In
addition, drying can be accelerated with a 10 minute drying phase with the
autoclave door open.
50/66