Setting Up Delivery Status; Surgical Motor In The Program - KaVo INTRAsurg 300 Instructions For Use Manual

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
PRG
PRG
▶ Press the program button for at least three seconds.
An acoustic signal confirms the entry.

5.4.6 Setting up delivery status

Note
All entries and changes made are deleted with this function.
Note
If the device was initialised, everything saved previously is deleted.
Therefore, for simplicity, everything that was saved should be noted down for easy
recovery.
▶ Turn device on while pressing the programme button.
After initialization, the device is ready for use in its condition at the time of de‐
livery.
Condition at the time of delivery: Programme
Step
Instrument
1
27:1
2
27:1
3
27:1
4
27:1
5
27:1
6
27:1
Condition at the time of delivery: Free application
Step
Instrument
27:1

5.4.7 Surgical motor in the program

Requirement
The mounted INTRA CL light, hand and contra-angle handpiece must be selected
on the display.
If the mounted handpiece or contra-angle handpiece is not selected, the device asks
for the corresponding input.
Speed (rpm) Torque
(Ncm)
11-1500
55
11-50
40
11-50
40
11-50
40
11-50
55
11-50
10
Speed (rpm) Torque
(Ncm)
11-1500
55
38/66
Pump
Direction of
motor rota‐
tion
4
->
4
->
4
->
4
->
0
-<
0
->
Pump
Direction of
motor rota‐
tion
4
->

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intrasurg 300 plus

Table of Contents