Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
Para-
me ter

5.3.2 Set the torque limitation

▶ Press the parameter button until the asterisk flashes at the torque display ③.
Note
The torque values ③ are only for KaVo handpieces and angle pieces that operate
properly.
The maximum value can be changed in the parameter menu by using the Plus and
Minus buttons.
Note
Once 90% of these set torque has been reached, a signal sounds. When the set
torque is exceeded during treatment, the motor stops. The foot pedal has to be
pressed again to start the motor.
Checking the handpieces and contra-angle handpieces:
Check the set transfer ratio, and let the motor and handpiece or contra-angle hand‐
piece run for 20 seconds at maximum speed. The torque ③ is an indicator for the
condition of the instrument.
Display of 0 = status is good
Display < 0.5 = improve maintenance
Display > 0.5 perform service

5.3.3 Preset the speed range

Note
Only the maximum value ① can be preset. Use the multifunctional foot control to
select the speed within the predetermined range by simultaneously pressing the
slide pedal and moving it to the side.