Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.2 Disinfection
W
Z
▶ Press the left button briefly.
You exit the rinsing menu.
Y

6.2 Disinfection

Note
The motor and motor hose cannot be thermodisinfected. After treating each patient,
disinfect the surfaces close to the patient that are contaminated by contact or aer‐
osols. Wipe disinfection should be used in all cases.

6.2.1 Automated disinfection

Not applicable for INTRAsurg device parts.

6.2.2 Manual disinfection

▶ Pull the insertion needle ③ out of the coolant reservoir ②.
▶ Open the lock ① and remove the hose ④.
▶ Remove all hoses marked gray including tips and surgical motor on the unit.
▶ Wipe-disinfect surfaces with a soft, disposable cloth and permissible disinfec‐
tants. Make sure that the device is completely wet. Let the disinfectant act for
the prescribed time. Dry the surface.
49/66