Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.5 Connect the surgery motor and coolant reservoir
CAUTION
Tipping danger from excessively heavy coolant reservoir
Malfunctions
▶ Do not use coolant reservoirs that can hold more than 1 litre!
▶ Ensure that the device is standing securely.
Insert a sterile ventilation needle into glass bottles. The flow will otherwise be re‐
duced if coolant is withdrawn for a long period. Check the level of the coolant res‐
ervoir several times after waiting a while. The system must be leak-free. No liquid
may escape from the coolant hose or the side of the connections.
Note
The saline solution corrodes all the metal parts. Immediately remove residue.
▶ Fasten the self-adhesive cable clip ⑦ Mat. no. 1.001.3348 to the INTRAsurg
300/300 plus (for the location, see the illustration).
▶ Insert the coolant hose ④ into the cable clip and close.
Note
Always monitor the level of the coolant reservoir. The device itself does not monitor
the level. Use transparent coolant reservoirs.
If the product is used on potentially infectious patients, use disposable products for
supplying coolant and thorough infection protection. KaVo recommends using a
sterile disposable protective hose over the motor and hose (approx. 1.2 m) and a
transparent disposable protective film over the device. The control buttons and dis‐
play must always be identifiable for operation.