Instructions for use SURGmatic S11 L – 1.009.1010, SURGmatic S11 C – 1.009.1005
5 Operation | 5.4 Remove the bur
Bur shaft slips inside the chuck due to excessive speed of the bur or
abrupt engagement of the bur.
Property damage to bur shaft and chuck system, reduction of the service life of
bur and chuck system.
4 Do not operate the bur at a higher speed than recommended by the manu-
facturer.
4 Rotate the clamping ring sleeve in the direction of the arrow to the bur stop
and insert the bur in the chuck.
4 Turn the clamping ring back into its initial position.
4 Make sure that the bur is seated securely by pulling on it.

5.4 Remove the bur

Rotating bur.
Cuts, infection and burn injury.
4 Do not touch the bur while it is rotating.
4 Remove the bur from the handpiece after treatment to avoid injury and in-
fection during storage.
4 After the bur has come to a standstill, turn the clamping ring in the direction
of the arrow to the bur stop and remove the bur.
4 Turn the clamping ring back into its initial position.

5.5 Modification for contra-angle burs

NOTE
The handpiece must be converted if you wish to use contra-angle burs.
4 Open the handpiece chuck.
4 Insert the enclosed bur stop in the chuck.
4 Press the contra-angle bur onto the bur stop, close the clamping ring and
make sure that it is firmly seated.
4 Use the enclosed hook to remove the bur stop.