Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.5 Connect the surgery motor and coolant reservoir
WARNING
Electric shock due to coolant ingress
Injuries
▶ Rotate the bottle holder away from the device to ensure that no leaking coolant
drips on the device or the electrical connection.

4.5 Connect the surgery motor and coolant reservoir

Note
The delivered parts are not sterile. Sterilise them before the first patient treatment.
All parts conducting liquids must be sterilised.
▶ Plug in the surgery motor ⑥ at these surgery hose coupling and tighten with a
union nut ⑤.
Follow the separate instructions for use for the motor.
▶ Place the surgery motor on the motor holder ①.
Note
The kink-protection part on the coolant hose inlet may not be twisted. Twisting the
kink protection can cause defects in the house and lead to a restriction of coolant
flow and causes the coolant to drip afterwards.