Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
Note
Always save the selected setting with enter button when leaving the menu. Other‐
wise, the motor will not start (safety circuit).
Examples of saving values of handpieces and contra-angle handpie‐
ces
Note
Default values for the respective transfer ratios can be selected with the Plus or
Minus buttons.
Indication
Transfer ra‐
tio
Grinding,
1:2
cutting
Grinding,
1:1
Drilling,
cutting
Grinding,
12:1
Drilling,
Thread,
Implant
Grinding,
16:1
Drilling,
Thread,
Implant
Grinding,
27:1
Drilling,
Thread,
Implant
Grinding,
36:1
Drilling,
Thread,
Implant
Note
The listed indications are only examples. To prevent unnecessary risk, observe the
guideline speeds given by the manufacturer of the rotating instruments. KaVo rec‐
ommends setting a speed range of 11-25/min or 11-35/min for screwing tasks.
Combined
unit
Head
Reducing
shank
3555
--
1:2
3,610 N
--
1:1
3624 N
C 3
4:1
3:1
3624 N
67 RIC
4:1
4:1
C 09
C 3
9:1
3:1
C 09
67 RIC
9:1
4:1
29/66
Speed
Maximum
range
torque
rpm
Ncm
600-80.000 2,7
300-40.000 5,5
25-3.300
40
18-2.500
45
11-1.500
55
9-1.100
55