KaVo INTRAsurg 300 Instructions For Use Manual page 43

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
▶ Place the INTRA CL light, hand and angle handpiece on the INTRA LUX SL 550
motor and turn it until the catch ① audibly locks in place.
The hand and angle piece is recognised, and the corresponding saved parameters
are shown on the device display ②.
CAUTION
Incorrect input values.
Danger of injury.
▶ The input values must be checked before each use.
These parameters are immediately activated in free use.
The recognised handpiece or angle piece must be selected in the program. If it is
not selected, the device asks for the corresponding input. Three acoustic warning
signals sound, and the read-in instrument flashes on the device display with a ques‐
tion mark, for example: ?27:1?
Note
For safety reasons, the motor does not start (safety circuit) since the input value
on the display always has priority.
▶ Check the parameters. Values displayed by the device can be saved by pressing
the ENTER button.
The data of the mounted INTRA CL light, hand and angle piece are only saved in
this step. The motor starts with these released parameters.
Note
This must be done for the saved parameters to be reliably maintained and contin‐
uously followed.
41/66