Exchange The Pump Hose - KaVo INTRAsurg 300 Instructions For Use Manual

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.1 Exchange the pump hose

7.1 Exchange the pump hose

Note
The parts delivered for exchange are not sterile. Rise and sterilize them before the
first treatment.
See also: 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060, Page
50
Note
If the pump hose becomes brown after extended use, it must be exchanged.
▶ Make sure that all lines containing saline solution are completely empty.
▶ To drain the lines, remove the insertion needle ③ from the saline solution res‐
ervoir ② and activate the pump by using the multifunctional foot control until the
hoses are free of saline solution. Remove the container and store it so that it
does not drain.
▶ Turn off the INTRAsurg 300/300 plus.
CAUTION
Running, open hose pump.
Injury hazard
▶ Turn off the device before opening the hose pump.
▶ Lift the lock and remove the pump hose.
Insert the new pump hose Mat. no. 1.001.3347.