Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
▶ Take the roller out of the instrument arm and then place the hose in the
germ reduction attachment.
▶ Pull-off the swing-arm in an upward direction.
▶ Clean the swing-arm and roller under running water.
▶ Wipe the instrument arm S table with disinfectant.
Sterilisation of instrument arm S table
Damage caused by sterilisation.
▶ Do not sterilise the instrument arm S table.
2.4.4 Setting up the 3-function and multifunction
handpiece
Note
Wet the O-rings of the gripping sleeves after sterilisation using silicone
grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no.
0.411.7510). After replacing the cannula, the air channel must be blown out
before the start of treatment to ensure that the air exiting during treatment
is dry.
Cleaning the triple-function and multifunctional
handpiece
Automatic cleaning
Not applicable.
Damage caused by improper cleaning/disinfection.
▶ Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶ Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶ Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.
NOTICE
NOTICE