Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
Para-
me ter
Para-
me ter
CAUTION
Excessive speed.
Injuries.
▶ In applications where the speed is critical, move the hinged pedal to the left
position (minimum speed) before the motor starts and move up to the desired
speed.
▶ Set the speed to a safe maximum for the application.

5.3 Free use

5.3.1 Parameters menu

The following values can be changed in the parameter menu:
▪ Transfer ratio of the handpiece and contra-angle handpiece
▪ max. speed
▪ Direction of motor rotation
▪ Coolant pump setting
▪ max. torque
▶ Press the parameter button.
A star ① flashes next to the alterable parameter.
▶ Select the desired parameter by pressing the parameter button several times.
An asterisk * flashes next to the alterable parameter.
▶ Set the parameters to the desired value with the plus or minus key.
▶ Press the Enter key.
The set values are saved.
28/66