Troubleshooting - KaVo INTRAsurg 300 Instructions For Use Manual

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
Malfunction
Nothing works.
No coolant in the instrument.
The coolant in the instrument is in‐
sufficient.
Coolant is dripping

7 Troubleshooting

Note
If your problem cannot be rectified after reading the following troubleshooting doc‐
umentation, please contact a KaVo-trained Service Technician.
In case of a problem, a fault number is displayed with: "warning xx" or "remark xx"
Fault numbers 1 to 67 are possible.
Note
Confirm all other error messages with the ENTER button. If the error message does
not disappear or the error is reported again, please notify customer service.
Please always write down what you want or the service you want performed when
you send in the device. The settings must be rechecked after the device is returned.
Cause
Fuse blown.
The unit must be switched off.
Both ends of the power line are not
plugged in.
The wrong handpiece or angle
piece is mounted or set on the de‐
vice.
Unknown.
Coolant flow not preselected.
The pump hose is sticky or crusty. ▶ Exchange the pump hose.
The spray nozzles are crusty or
soiled.
The pump hose is sticky or crusty. ▶ Exchange the pump hose.
Glass bottle with saline solution is
not ventilated.
Clogged coolant line or restricted
flow
Remedy
▶ Check and if necessary replace
▶ Turn on the mains switch on the
▶ Plug in the mains power line:
▶ Adjust the right handpiece or an‐
▶ Turn the machine off and on.
▶ Set the coolant flow.
See also: 5.3.4 Set the coolant flow,
Page 31
See also: 7.1 Exchange the pump
hose, Page 56
▶ Clean the spray nozzle with the
See also: 7.1 Exchange the pump
hose, Page 56
▶ Ventilate the glass bottle.
▶ Check if the coolant hose and
53/66
the fuse.
back of the unit.
gle piece on the device.
nozzle needle.
connecting nipple are blocked.

Hide quick links: