6 Preparation methods DIN EN ISO 17664 | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
▶ Wind all hoses previously cleaned and disinfected that are marked grey including
tips and the surgical motor on the motor holder ⑥.
Seal the motor holder, handpieces and contra-angle handpieces in sterile bags
or place them on a surgical tray.

6.3.2 Sterilisation

▶ Sterilise in saturated steam at 135±1℃, gravity, for at least 4 minutes.
Note
Let the sterilised materials cool and try to the ambient temperature before using
them again.

6.3.3 Storage

Observe all necessary measures for hygiene when storing sterile goods. Store pro‐
tected from dust, release with identification on the packaging. Monitor storage
length.