6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning

6.1 Cleaning

Always clean directly after using the medical device to prevent contaminants from
hardening. Saline solution contains sodium chloride which corrodes metal. Quickly
remove all residue to prevent surface damage.

6.1.1 Machine cleaning

Not applicable for INTRAsurg device parts.
CAUTION
Damage due to liquids
Faults in electric components.
▶ Do not machine-clean the INTRAsurg device.

6.1.2 Manual cleaning

▶ Wipe off the supply unit (all visible surfaces), motor hose, motor holder, bottle
container, foot control surfaces and the connecting lines with a damp disposable
cloth. Do not wipe the cloth on edges and press so that liquid will not penetrate.
Remove all residue and dry the device.
Note
After uses that run the risk of infection, rinse the coolant lines of the handpieces
and contra-angle handpieces with a disinfectant (approx. 20 ml) with a disposable
syringe. Avoid damage. Rinse the coolant lines with demineralised water, then
drain and dry them.

6.1.3 Rinse the hoses

Preparation for product decontamination:
Directly after each use, rinse all the hoses to prevent malfunctions from the crys‐
tallisation of the saline solution. All of the hoses must be completely emptied. Then
sterilise them and observe the hygienic requirements for storage and maintaining
sterility.