Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning
W
Z
▶ Remove the pump hose needle from the coolant reservoir. Immerse the pump
hose needle in a clean container with demineralised water (50-60 ml). Place the
spray hose from the instrument in an empty container or sink.
Note
Rinse until all the hoses are drained. Do not immerse the water outlet into the liquid.
Observe the necessary measures for infection protection.
▶ Press left button 3 sec. The rinsing function is selected.
See also: 5.3.7 Rinsing function, Page 33
Y
▶ The rinsing function is shown in the display.
▶ Hold down slide pedal and slide it to the right until the stop. The hose pump runs
at the max. rate.
▶ Ensure that the water jet is in the right direction and that it is the right amount.
48/66