Request Maximum Torque - KaVo INTRAsurg 300 Instructions For Use Manual

Surgical implant rotary system non optic unit
Table of Contents

Advertisement

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
Para-
me ter
if the motor is turned on without the corresponding entry, three acoustic signals
sound, and the read-in instrument flashes on the device display with question marks,
for example: ?27:1? .
Note
The motor will not start (safety circuit).
To start the motor, either the selected handpiece or contra-angle handpiece must
be mounted, or the mounted handpiece or contra-angle handpiece must be saved
using the ENTER button.
Save the light, hand and contra-angle handpieces (without the moun‐
ted and peace and contra-angle handpiece)
▶ Press the parameter button until the asterisk flashes at the transfer ratio.
▶ Set the correct parameters with the Plus or Minus button.
▶ Save the set values with the enter button.
The data are now saved in this step.

5.4.8 Request maximum torque

The maximum torque is saved when you advance a step.
When the motor is started in the next step, the highest torque is always saved.
▶ Press the enter button after closing the program.
In each step ①, the max. torque ② is shown in the display.