Operating Instruction EXTRAmatic 9040
|
5 Operation
5.3 Utility programs
You can set the memory time to between 0 and 20 seconds.
This means that, for a set amount of time, the extractor unit 'remembers' the ex‐
traction level most recently chosen at a particular workstation.
If the workstation is switched back on during this memory time, extraction continues
at the most recently selected extraction level; if it is not switched back on during this
time, extraction commences at the level configured in the "User system" utility pro‐
gram.
Factory setting: 5 seconds
▶ Press ENTER.
This saves the new value that has been set.

5.3.16 "SW version" utility program (2700)

▶ Using the up or down button, select the "Software version" utility program.
The software version for this extractor unit will then be displayed (e.g. "SW ver‐
sion 2.0").

5.3.17 "Regular filter cleansing time" utility program (2750)

▶ Using the the up or down button, select the "Filter cleanse time" utility program
and choose ENTER.
▶ Using the up or down button, set the required filter cleaning interval time (bet‐
ween 5 and 120 minutes).
This is the interval between the regular filter cleaning.
Factory setting: 60 minutes
▶ Press ENTER.
This saves the new value that has been set.