KaVo Primus 1058 Life Care Instructions
Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

Quick Links

Care instructions
Primus 1058 Life

Advertisement

Table of Contents
loading

Summary of Contents for KaVo Primus 1058 Life

  • Page 1 Care instructions Primus 1058 Life...
  • Page 2 Manufacturer: Distributed by: Kaltenbach & Voigt GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach www.kavo.com Phone +49 (0) 7351 56-0 Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Target group ...................... 4 2 Setup methods according to DIN EN ISO 17664..............5 2.1 References........................5 2.2 General instructions for the reprocessing of KaVo treatment units........6 2.2.1 Reprocessing Overview Primus 1058 Life............... 7 2.3 Reprocessing of the surfaces of treatment unit / upholstery ..........8 2.4 Preparation of the dentist unit ..................
  • Page 4: User Instructions

    1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. Note Comply with the Instructions for Use of the Primus 1058 Life treatment unit! 1.1.1 Abbreviations Explanation brevi- ation Instructions for use...
  • Page 5: Setup Methods According To Din En Iso 17664

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.1 References 2 Setup methods according to DIN EN ISO 17664 2.1 References The reprocessing consists of the following basic steps: ▪ Cleaning and disinfection ( manually or automatically) ▪...
  • Page 6: General Instructions For The Reprocessing Of Kavo Treatment Units

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units 2.2 General instructions for the reprocessing of KaVo treatment units Please comply with the general reprocessing instructions at all times. Detailed product-specific reprocessing instructions are provided subsequently.
  • Page 7: Reprocessing Overview Primus 1058 Life

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units Manual drying ▶ Supply dry, clean and uncontaminated compressed air according to ISO 7494-2 only.
  • Page 8: Reprocessing Of The Surfaces Of Treatment Unit / Upholstery

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit / upholstery Reprocessing items/medical Cleaning Disinfection Sterilisa- devices tion Connection parts of the suction hose Patient element...
  • Page 9: Preparation Of The Dentist Unit

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit NOTICE Damage caused by spray disinfection. Disinfectant gets into cracks and loosens the paint. ▶ Do not use spray disinfection.
  • Page 10 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Cleaning and disinfection of S-table handle and holder pad/anti-slip mat ▶ Press the push button and pull off the handle.
  • Page 11: Reprocessing Of The Instrument Holder Of The Tm Table

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Sterilisation of S-table handle and holder pad/anti-slip NOTICE Damage caused by sterilisation. ▶ Do not sterilise the white standard handle.
  • Page 12 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Place the instrument hoses in the corresponding places of the germ reduc- tion attachment during the cleaning and germ reduction of the instrument holder.
  • Page 13: Reprocessing Of The Instrument Arm S Table

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit 2.4.3 Reprocessing of the instrument arm S table Note The instrument arm S table (Mat. no. 1.007.4756) can be removed to facil- itate cleaning.
  • Page 14: Setting Up The 3-Function And Multifunction Handpiece

    Note Wet the O-rings of the gripping sleeves after sterilisation using silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replacing the cannula, the air channel must be blown out before the start of treatment to ensure that the air exiting during treatment is dry.
  • Page 15 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Manual cleaning Note Combine manual cleaning with disinfection. Cleaning the interior ▶ Leave the gripping sleeve and cannula on the instrument, and activate the air and water for 60 seconds.
  • Page 16 Requirement The multifunctional handpiece has been reprocessed. ▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510) to the O-rings of the joint between the gripping sleeve and cannula. Use cotton buds for this purpose.
  • Page 17: Preparation Of The Instrument Hoses

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Pull the gripping sleeve off the valve body. Sterilization in a steam sterilizer in compliance with DIN EN 13060 ▶...
  • Page 18: Servicing The Turbine Return Air Filter

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit 2.4.6 Servicing the turbine return air filter Cleaning the turbine return air filter Note The liquid collector must be checked weekly.
  • Page 19 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Clean the surface with a soft cloth and mild liquid cleaning agent. Disinfection Exterior disinfection by hand WARNING Highly flammable gases-.
  • Page 20: Reprocessing Of The Assistant Unit

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant unit ▪ Microcide AF Liquid (Schülke & Mayr) ▪ FD 322 (Dürr Dental) ▪ Incidin Liquid (Ecolab) Note The exterior of the camera and camera hose can be disinfected with chemical disinfectants.
  • Page 21: Reprocessing Of Sieve Inserts And Suction Hoses

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit 2.5.3 Reprocessing of sieve inserts and suction hoses Note Check the sieve inserts daily and after each surgery. The sieve inserts must be replaced at the latest when the suction decreases (Mat.
  • Page 22 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit Cleaning and disinfection of the tumbler holder and filler Note When assembling, make sure that the O-ring is present on the tumbler filler.
  • Page 23: Preparation Of The Spittoon

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit 2.6.3 Preparation of the spittoon NOTICE Water in the device. Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl flush"...
  • Page 24: Manual Reprocessing Of The Water-Conducting System

    (Mat. no. 0.489.3451). ▶ If the Water bottle kit is used with the enclosed dosing attachment (Mat. no. 1.002.0287), add the proper amount of KaVo OXYGENAL 6 (Mat. no. 0.489.3451) with each filling. For the correct amount, please refer to the instructions of the dosing attachment for water germ reduction.
  • Page 25: Rinsing Programme

    During normal operation of the treatment centre, the continuous germ reduc- tion function secures continuous germ reduction in the water conducting sys- tems due to the automatic dosing of the germ reduction liquid, KaVo OXY- GENAL 6. Rinsing programme and intensive germ reduction facilitate the rinsing and germ reduction of the water-supplying system.
  • Page 26 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys- Place the instruments on the sterilisation attachment with patient part assembly kit / without patient part assembly kit The cannula of the triple-function and multifunctional handpieces must be re- moved for the rinsing program and the intensive germ reduction.
  • Page 27 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys- ▶ Insert the triple-function or multifunctional handpieces without the cannula into the clamping device of the germ reduction attachment. Make sure that the "W"...
  • Page 28: Intensive Germ Reduction

    During intensive germ reduc- tion, the afferent water ducts are automatically filled with a higher concentra- tion of KaVo OXYGENAL 6. The intensive germ reduction program lasts ca. 45 minutes. The increased hydrogen peroxide concentration acts in the system for at least 30 minutes.
  • Page 29 If an audible signal is issued every 10 seconds and the bottom LED of the "In- tensive germ reduction" button flashes, the KaVo OXYGENAL 6 must be replen- ished. If a beep is issued during the filling process (10 times at a one-second interval), the container is full.
  • Page 30 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys- Step Description Dwell time (acoustic signal, switch off is possible) Duration: 30 minutes Rinsing of the tumbler...
  • Page 31 The unit is now ready for use. Additional intensive germ reduction during period of non-use KaVo recommends carrying out an additional intensive germ reduction at uni- versities during semester breaks. This flushes the parts of the unit and replaces the germ reduction liquid.
  • Page 32: Reprocessing And Automatic Hygiene Functions Of The Suction System

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suc- tion system 2.9 Reprocessing and automatic hygiene functions of the suction system 2.9.1 Reprocessing of the suction system including...
  • Page 33: Preparation Of The Amalgam Separator

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.10 Preparation of the amalgam separator ter (limited rinsing). The water is aspirated from the Aquamat through the filter housing to the separation system. The spittoon bowl rinsing is alternately turned on and off.
  • Page 34: Reprocessing Kavo Screen One / Kavo Screen Hd

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.11 Reprocessing KaVo Screen One / KaVo Screen HD 2.11 Reprocessing KaVo Screen One / KaVo Screen HD Cleaning and disinfection NOTICE Non-colour-fast textiles. Non-colour-fast textiles can cause permanent discoloration.
  • Page 35 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.12 Reprocessing of the service table 1568 Note The handle on the service table can be taken off to improve the cleaning properties. ▶ Push the buttons ① on the bottom of the handle, and pull the handle off in the direction of the arrow.
  • Page 36: Replenish And Check The Disinfectant

    10 seconds. Note Replenish KaVo OXYGENAL 6 only if there is a beep every 10 seconds. The signal can be turned off as follows: Press the "Intensive germ reduction" but- ton briefly. The next time the treatment centre is turned on, the beep is is- sued again.
  • Page 37: Replenish Oxygenal 6 When Using The Water Bottle

    Care instructions Primus 1058 Life 3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6 Note Do not fill anything into the central opening (tumbler drain). ▶ Quit replenishing immediately if you hear an acoustic signal being issued every second (10 sounds).
  • Page 38 ▶ Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility. ▶ Guide the tube through the orifice of the coarse sieve of the KaVo OXY- GENAL 6 bottle. ▶ Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL 6 bottle.
  • Page 39 Care instructions Primus 1058 Life 3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6 ▶ Then fill-up the water bottle with tap water. Note Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended periods of non-use after condensation of the water.
  • Page 40: Check The Oxygenal 6 Concentration

    Care instructions Primus 1058 Life 3 Replenish and check the disinfectant | 3.2 Check the OXYGENAL 6 concentration 3.2 Check the OXYGENAL 6 concentration ▶ Test the standard concentration 0.02% H (200ppm): Fill a measuring cup (40 ml) with 39 ml water and 1 ml treated water.
  • Page 41 Care instructions Primus 1058 Life 3 Replenish and check the disinfectant | 3.2 Check the OXYGENAL 6 concentration ð Compare the colours and read the value: 5 mg/l is shown for a concentra- tion of 200ppm. Other concentrations can be tested using the following table:...