Analyzer Options Available; Option Ld5, High Stability Frequency Reference; Option 002, Harmonic Mode; Option 006,6 Ghz Operation - HP 8753E User Manual

Network analyzer
Table of Contents

Advertisement

Analyzer Options Available

Option lD6, High Stability Frequency Reference
Option lD5 offers kO.05 ppm temperature stability from 0 to 60 OC (referenced to 25 "C).

Option 002, Harmonic Mode

Provides measurement of second or third harmonics of the test device's fundamental output
signal. Frequency and power sweep are supported in this mode. Harmonic frequencies can be
measured up to the maximum frequency of the receiver. However, the fundamental frequency
may not be lower than 16 MHz.
Option 006, 6 GHz Operation
Option 006 extends the maximum source and receiver frequency of the analyzer to 6 GHz.

Option 010, Time Domain

This option displays the time domain response of a network by computing the inverse Fourier
transform of the frequency domain response. It shows the response of a test device as a
function of time or distance. Displaying the reflection coefficient of a network versus time
determines the magnitude and location of each discontinuity. Displaying the transmission
coefficient of a network versus time determines the characteristics of individual transmission
paths. Time domain operation retains all accuracy inherent with the correction that is active in
the frequency domain. The time domain capability is useful for the design and characterization
of such devices as SAW iilters, SAW delay lines, RF cables, and RF antennas.

Option 011, Receiver Configuration

Option 011 allows front panel access to the R, A, and B samplers and receivers. The transfer
switch, couplers, and bias tees have been removed. Therefore, external accessories are
required to make most measurements.
Option 076, 768 Impedance
offers 75 ohm impedance bridges with type-N test port connectors.
Option 075

Option lDT, Delete Display

This option removes the built-in flat panel display, allowing measurement results to be viewed
with an external monitor only.

Option lCM, Rack Mount Flange Kit Without Handles

Option 1CM is a rack mount kit containing a pair of flanges and the necessary hardware to
mount the instrument, with handles detached, in an equipment rack with 482.6 mm (19 inches)
horizontal spacing.
HP 8753E Description and Options

Hide quick links:

Advertisement

Table of Contents
loading