4. Connect adapter A3 to adapter Al on port 1. (See F'igure 5-13.)
5. Perform the 2-port error correction using calibration standards appropriate for the
connector type at port 2.
6.
Save the results to disk. Name the fSle "PORTB."
7. Determine the electrical delay of adapter A3 by performing steps 1 through 7 of "Modify
the Cal Kit Thru Definition."
Remove the Adapter
When the two sets of error correction fles have been created (now referred to as "cal sets"),
the adapter may be removed.
, ~~~~~.~~~~~~(This provides a quick reference guide to using *e adapter
removal technique.)
. ~~~~~~~:
n ~~~~~~~~~
. .: : : : . .: : . . .
. .i . : . .
.
.
.
.
.
N E T W O R K A N A L Y Z E R
Figure 5-13. Two-Port Cd
REFERENCE
PORT 2
Set 2
Optimizing Measurement Results 543