Environmental Characteristics

General Conditions
EMC characteristics: emissions, CISPR Publication 11; immunity, IEC 801-2/3/4, level 2.
Electrostatic discharge (ESD): must be eliminated by use of static-safe work procedures and
an anti-static bench mat (such as HP 92175T).
Dust: the environment should be as dust-free as possible.
Enclosure protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IP code 2 0, according to IEC 529
Operating Conditions
Operating temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . o" to55OC
Error-Corrected temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . fl "C of calibration temperature
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5% to 95% at 40 "C (non-condensing)
Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 4500 meters (15,000 feet)
Non-Operating Storage Conditions
Temperature.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -40 "C to +70 OC
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 90% relative at +65 OC (non-condensing)
to 15,240 meters (50,000 feet)