3. To create a test sequence, enter the parameters for the measurement that you wish to make.
For this example, a SAW filter measurement is set up with the following parameters:
The above keystrokes will create a displayed list as shown:
Start of Sequence
RECALLPRSTSTATE
LOG MAG
CENTER
134 M/u
SPAN
50 M/u
SCALE/DIV
AUTO SCALE
4. To complete the sequence creation, press:
When you create a sequence, the analyzer stores it in volatile memory where
Caution
it will be lost if you switch off the instrument power (except for sequence #6
which is stored in the analyzer non-volatile memory). However, you may store
sequences to a floppy disk.

Running a Sequence

To run a stored test sequence, press:
or, press:

Stopping a Sequence

To stop a sequence before it has finished, press m.