Band Switch Points - HP 8753E User Manual

Network analyzer
Table of Contents

Advertisement

3. Then switch to stepped list mode:
If there is no difference between the measurements in either list mode, then use the swept
n
If the memory trace indicates that there is more attenuation in swept list mode, it may be
n
due to IF delay. You can usually remedy this problem by increasing the sweep time.
Note
IF bandwidths of 30 to 10 Hz cause the sweep (or that segment of the sweep)
to be stepped, thus eliminating IF delay.
The hardware of the network analyzer sweeps the frequency range in separate bands, where
switching from band to band takes time. Modify the frequency span to ehminate as many band
switches as possible while maintaining measurement integrity. Refer to the following table to
identify the analyzer's band switch points:
Auto sweep time mode is the default mode (the preset mode). This mode maintains the fastest
sweep speed possible for the current measurement settings.
n press IMenu) ~~~~,'~~~ @ @J to re-enter the auto mode.
.......
...... .... ...... ......
.................. ............. .............................
Band
I
Frequency Span
3.3 MHz to 16 MHz
16 MHz to 31 MHz
31 MHz to 61 MHz
61 MHz to 121 MHz
121 MHz to 178 MHz
178 MHz to 296 MHz
296 MHz to 536 MHz
536 MHz to 893 MHz
893 MHz to 1.607 GHz
1.607 GHz to 3 GHz
3 GHz to 4.95 GHz
4.95 GHz to 6 GHz
Optimizing Measurement Results 5-51