Caution
To prevent connector damage, use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.

Connections for the Second Portion of Conversion Compression Measurement

8. To set the frequency offset mode Lo frequency, press:
9. To select the converter type, press:
10. 'Ib select a low-side Lo measurement conflguration, press:
In this low-side Lo, up converter measurement, the analyzer source frequency is offset lower
than the receiver frequency. The analyzer source frequency can be determined from the
following equation:
receiver frequency (800 MHz) - Lo frequency (600 MHz) = 200 MHz
330 Making Mixer Measurements
NETWORK ANALYZER
Figure 3-21.
MIXER
UNDER TEST
3 dB
EXTERNAL
LO SOURCE